OpaCem is a polymethyl methacrylate bone cement developed for vertebral augmentation. The powder contains 60% zirconium dioxide, a radiological contrast agent, so the cement is clearly visible under fluoroscopy and the surgeon can control exactly where it goes in the vertebral body.
The viscosity is suited to injection in vertebroplasty and kyphoplasty. Mixing takes 30 seconds and the application time is long: at 23 °C, almost 8 minutes. In the manufacturer’s radiopacity test, OpaCem matched 4.6 mm of aluminium, compared with 2.2 mm for the previous version, NTCem Spine, with 30% zirconium dioxide.

| Powder (20 g) | Polymethyl methacrylate 7.70 g; zirconium dioxide 12.00 g; benzoyl peroxide 0.30 g |
|---|---|
| Liquid (8.5 g) | Methyl methacrylate (monomer) 8.436 g; N,N-dimethyl-p-toluidine 0.064 g; hydroquinone 150 ppm |
| Times at 19 °C | Mixing 30 s, waiting 170 s, application 610 s, hardening 120 s |
| Times at 23 °C | Mixing 30 s, waiting 130 s, application 460 s, hardening 80 s |
| Times at 26 °C | Mixing 30 s, waiting 90 s, application 350 s, hardening 70 s |
| Conditions for times | Components stored for 2 hours at 20 °C before use |
| Compressive strength | 90-100 MPa |
| Bending strength; modulus | 60-70 MPa; 4,909 MPa |
| Application time; setting time | 8-10 min; 12-14 min (ISO 5833) |
| Exothermic temperature | 60-80 °C |
| Intrusion; radiopacity | ≥ 2 mm; 4.6 mm Al |
| Order number | 4210BG0002 |
| Marking | CE 2195 |

It is the radiological contrast agent. The powder is 60% zirconium dioxide, so the cement is clearly visible under fluoroscopy and the surgeon can control where it goes in the vertebral body. In the manufacturer’s test, radiopacity was equivalent to 4.6 mm of aluminium.
It depends on temperature. At 23 °C: mixing 30 s, waiting 130 s, application 460 s and hardening 80 s; at 19 °C the times are longer, and at 26 °C shorter.
In vertebroplasty and kyphoplasty, for severely painful osteoporotic vertebral compression fractures, painful vertebral body tumours (metastases or myeloma) and symptomatic vertebral haemangioma.
Compressive strength of 90-100 MPa, bending strength of 60-70 MPa and an exothermic temperature of 60-80 °C, tested to ISO 5833.
Product
OpaCem
Meta Biomed
Fill in the form and we will get back to you shortly, usually within the same working day.

A complete solution for unilateral biportal endoscopic spine surgery (UBE): 4K endoscopic tower, Plasma and RF electrodes, dedicated instruments and an endoscopic drill.

A complete sterile set for biportal and monoportal spinal endoscopy: drapes, camera cover, fluid collection and the small instruments needed, in a single pack.

A modular anterior cervical fusion (ACDF) system that combines a TiPEEK interbody cage with an anterior plate: 4 plate types, rigid or variable fixation, 8 configurations.