EnSight Biportal
A complete solution for unilateral biportal endoscopic spine surgery (UBE): 4K endoscopic tower, Plasma and RF electrodes, dedicated instruments and an endoscopic drill.

The self-locking polymer cerclage is a braided polyester (PET) cable, an alternative to the metal cable traditionally used for cerclage of long bones in traumatology and orthopaedics. The manufacturer’s technique describes cerclage of a femoral fracture, and the steps are the same for other long bones. The surgeon may use additional osteosynthesis material with the cerclage, such as a plate, for reduction and healing.
The implant has an adjustable loop, a titanium connector, a fixation loop, a splice loop and a semi-rigid end for the cable passer. The cable is passed around the bone with a Ø 41 or 52 mm cable passer, the fixation loop is clipped into the connector, and tension is applied with the dedicated tensioner. The splice holds the tension after three equal successive tightenings; a 2.3 Nm torque-limiting handle can be fitted to the tensioner.
The cerclage can be positioned before fracture reduction but is tightened only after reduction. When used with a plate, the cable is preferably passed over the plate and tensioned after the screws are fixed. Once all cerclages are tensioned and checked, the tensioning end is cut 2 cm from the splice loop; from then on the tension can no longer be adjusted.

| Implant | Braided PET cable (OCBTRCER1U) |
|---|---|
| Tensioner | OCBANCER1U |
| Torque-limiting handle | 2.3 Nm (0110023060) |
| Cable passers | Small, Ø 41 mm (0105010010); large, Ø 52 mm (0105030009) |
| Trimming the end | 2 cm from the splice loop, after all cerclages are tensioned |
| Classification | Implant Class IIb, CE 1639 (SGS Belgium NV); tensioner, cable passers and handle, Class I |
| Manufacturer | Cousin Biotech S.A.S., France (cable passers and handle: SAYAN TIBBI ALETLER) |


In long bone fractures, traumatic or in reconstructive surgery, and to reattach the greater trochanter in hip arthroplasty or other procedures with a trochanteric osteotomy.
Through the self-locking splice loop: after three equal successive tightenings with the dedicated tensioner, the system holds the chosen tension. A 2.3 Nm torque-limiting handle can be fitted to the tensioner.
Yes. The cable is preferably passed over the plate and tensioned after the screws are fixed. In exceptional cases the cerclage can pass between bone and plate, provided the plate is smooth, without sharp edges, and the cable does not touch the screws.
Product
Self-locking polymer cerclage
Cousin Surgery
Fill in the form and we will get back to you shortly, usually within the same working day.

A complete solution for unilateral biportal endoscopic spine surgery (UBE): 4K endoscopic tower, Plasma and RF electrodes, dedicated instruments and an endoscopic drill.

A complete sterile set for biportal and monoportal spinal endoscopy: drapes, camera cover, fluid collection and the small instruments needed, in a single pack.

A modular anterior cervical fusion (ACDF) system that combines a TiPEEK interbody cage with an anterior plate: 4 plate types, rigid or variable fixation, 8 configurations.