title: "MySpine patient-specific 3D-printed spine guides"
description: "MySpine by Medacta: patient-specific 3D-printed guides for posterior fixation screws from C2 to the sacrum and ilium, planned on a low-dose CT scan."
url: https://rondamedtech.ro/en/producatori/medacta/myspine/
lang: en
---

1.  [Home](https://rondamedtech.ro/en/)
2.  [Manufacturers](https://rondamedtech.ro/en/producatori/)
3.  [Medacta](https://rondamedtech.ro/en/producatori/medacta/)
4.  MySpine

[![Medacta](https://rondamedtech.ro/logos/medacta.svg)](https://rondamedtech.ro/en/producatori/medacta/)

# MySpine

Patient-specific 3D-printed surgical guides that direct pilot-hole preparation for posterior fixation screws from C2 down to the sacrum and ilium, following a plan made by the surgeon on the patient’s CT scan.

[Neurosurgery](https://rondamedtech.ro/en/specialitati/neurochirurgie/) · [Orthopaedics and traumatology](https://rondamedtech.ro/en/specialitati/ortopedie/)

![MySpine guides on a spine model, with the 3D screw planning on screen](https://rondamedtech.ro/produse/medacta/myspine-coloana-planificare.webp)

## Description

MySpine turns the CT scan of the patient’s spine into a 3D preoperative plan and into anatomical drilling guides used once, for a single patient. In the MySpine Webplanner the surgeon sets the entry point, trajectory, diameter and length of each screw. The guide rests on the vertebra at several contact points on distinct anatomical landmarks, such as the spinous and transverse processes, and distal windows keep the operative field in view at every step.

The patient has a CT scan under the low-dose MySpine protocol, and the data are sent securely to Medacta, which reconstructs the anatomy in 3D. Once the surgeon has reviewed and approved the plan, the guides and plastic vertebral models are 3D printed. The surgeon receives them before surgery and can rehearse the placement of each guide on the model. Cases are managed entirely online, with no software to install, and for the first procedures the surgeon can ask for a Medacta proctoring surgeon through the M.O.R.E. programme.

The guides are designed for Medacta’s M.U.S.T. screw systems and their instruments. MySpine is part of the MySolutions ecosystem, alongside the NextAR platform, and the MySpine preoperative planning suite is also used by NextAR Spine. In a clinical study cited by the manufacturer (Van Royen et al., European Spine Journal, 2021), superimposing CT scans on the planned trajectories in thoracolumbar deformity surgery showed a mean deviation of 0.92 mm and 2.92°.

![MySpine thoracolumbar guide seated on the vertebral model](https://rondamedtech.ro/produse/medacta/myspine-ghid-toracolombar.webp)

## Indications

-   Posterior screw fixation of the cervical (C2-C7), thoracic and lumbar (T1-L5) spine
-   Posterior lumbar fusion along a midline cortical trajectory, from L1 to S1 (MySpine MC)
-   Posterior sacroiliac fixation, for example in long constructs with insufficient lumbosacral anchorage (MySpine S2AI and MySpine Anchor)
-   Adolescent and adult scoliosis, deformities that need long constructs and, according to the manufacturer, revision surgery

## Features

-   MySpine Cervical: bilateral or monolateral (right, left) guides for C2-C7 pedicle trajectories; the spinous process block can be open, open cranially, open caudally or closed
-   Monolateral cervical guides reduce muscle retraction
-   Thoracolumbar guides for the conventional trajectory, in a K-wire-based or drill-based configuration
-   Standard guides are intended for deformities and long constructs, Low Profile guides for the lower lumbar region and degenerative cases operated through a small incision
-   MySpine MC: entry points at the pars interarticularis and a midline, minimally invasive, muscle-sparing approach; 4-5 cm skin incision, according to the manufacturer
-   MySpine S2AI: the S2-alar-iliac trajectory crosses five cortical bone structures, and the medial entry point allows direct rod connection without additional connectors
-   Plastic model of each treated vertebra, supplied together with the guides
-   MySpine Webplanner: online 3D planning on 2D CT images and on the 3D model, with the case database available at any time
-   Single-use guides, with a sterile-supplied option; according to the manufacturer, the technology requires no capital investment

## Technical specifications

Technical specifications: MySpine

Imaging

CT scan under the low-dose MySpine protocol

Planning

MySpine Webplanner, online, no installation

Planned parameters

Entry point, trajectory, screw diameter and length

Manufacturing

3D printing: guides and vertebral models

Cervical

C2-C7; bilateral or monolateral guide

Thoracolumbar

T1-L5; Standard or Low Profile, K-wire- or drill-based

Lumbar, midline cortical trajectory

L1-S1 (MySpine MC)

Sacral

S1 and S2AI; Anchor

Screw systems

M.U.S.T. MINI (cervical), M.U.S.T. and M.U.S.T. MC (thoracolumbar), M.U.S.T. SI (sacroiliac)

Use

Single patient, single use

## Gallery

![Bilateral MySpine Cervical guide on a cervical vertebra](https://rondamedtech.ro/produse/medacta/myspine-cervical-bilateral.webp)

![Monolateral MySpine Cervical guide, right-side version](https://rondamedtech.ro/produse/medacta/myspine-cervical-monolateral.webp)

![MySpine MC guide above the lumbar vertebra it fits onto](https://rondamedtech.ro/produse/medacta/myspine-mc.webp)

![MySpine S1 and S2AI guide on the sacral model](https://rondamedtech.ro/produse/medacta/myspine-s2ai.webp)

![MySpine Anchor guide for posterior sacroiliac fixation](https://rondamedtech.ro/produse/medacta/myspine-anchor.webp)

![MySpine Webplanner on a laptop, tablet and phone, with the trajectories planned on the vertebra](https://rondamedtech.ro/produse/medacta/myspine-webplanner.webp)

## Documents

-   [MySpine brochurePDF · en · 1,8 MB](https://rondamedtech.ro/documente/medacta/myspine-brosura.pdf)
-   [MySpine Cervical brochurePDF · en · 1,5 MB](https://rondamedtech.ro/documente/medacta/myspine-cervical-brosura.pdf)
-   [MySpine S2AI brochurePDF · en · 533 KB](https://rondamedtech.ro/documente/medacta/myspine-s2ai-brosura.pdf)
-   [Adolescent scoliosis, patient informationPDF · en · 3 MB](https://rondamedtech.ro/documente/medacta/myspine-scolioza-adolescent-pacienti.pdf)
-   [Adult scoliosis, patient informationPDF · en · 1,1 MB](https://rondamedtech.ro/documente/medacta/myspine-scolioza-adult-pacienti.pdf)

## Frequently asked questions

### Who decides where the screws go?

The surgeon. In the MySpine Webplanner, on the 3D reconstruction of the CT scan, the surgeon chooses the entry point, trajectory, diameter and length of each screw. The guides are designed and printed only after the surgeon has approved the plan.

### What does the surgeon receive before surgery?

The patient-specific MySpine guides and a plastic model of each treated vertebra, both 3D printed. The model can be used to rehearse guide placement before the procedure.

### Which spinal segments is MySpine used on?

From C2 down to the sacrum and ilium. MySpine Cervical covers C2-C7, the thoracolumbar guides T1-L5, MySpine MC the L1-S1 segment on a midline cortical trajectory, and MySpine S2AI and MySpine Anchor posterior sacroiliac fixation.

### How does MySpine relate to NextAR Spine?

Both belong to Medacta’s MySolutions ecosystem. NextAR Spine uses the MySpine preoperative planning suite but guides the surgeon through augmented reality, whereas MySpine turns the plan into physical 3D-printed guides that sit on the vertebra.

## Quote request

Product

MySpine

Medacta

Fill in the form and we will get back to you shortly, usually within the same working day.

## Related products

Nexon Medical

### [EnSight Biportal](https://rondamedtech.ro/en/producatori/nexon-medical/ensight-biportal/)

A complete solution for unilateral biportal endoscopic spine surgery (UBE): 4K endoscopic tower, Plasma and RF electrodes, dedicated instruments and an endoscopic drill.

Product page

Nexon Medical

### [EnSight Drape Pack](https://rondamedtech.ro/en/producatori/nexon-medical/ensight-drape-pack/)

A complete sterile set for biportal and monoportal spinal endoscopy: drapes, camera cover, fluid collection and the small instruments needed, in a single pack.

Product page

Medacta

### [Mecta-C Stand Alone](https://rondamedtech.ro/en/producatori/medacta/mecta-c/)

A modular anterior cervical fusion (ACDF) system that combines a TiPEEK interbody cage with an anterior plate: 4 plate types, rigid or variable fixation, 8 configurations.

Product page

---
Quotes and information: https://rondamedtech.ro/en/contact/ · office@rondamedtech.ro · +40 727 221 112