---
title: "BDyn posterior dynamic stabilisation (T10-S1)"
description: "BDyn by Cousin Surgery: posterior dynamic stabilisation from T10 to S1 with a silicone cushion and PCU ring, plus hybrid rods for fusion constructs."
url: https://rondamedtech.ro/en/producatori/cousin-surgery/bdyn/
lang: en
---

1.  [Home](https://rondamedtech.ro/en/)
2.  [Manufacturers](https://rondamedtech.ro/en/producatori/)
3.  [Cousin Surgery](https://rondamedtech.ro/en/producatori/cousin-surgery/)
4.  BDyn

[![Cousin Surgery](https://rondamedtech.ro/logos/cousin-surgery.svg)](https://rondamedtech.ro/en/producatori/cousin-surgery/)

# BDyn

A spinal shock absorber (flexible spinal rod) for posterior dynamic stabilisation of the T10-S1 segments that preserves segmental motion and can combine fusion with dynamic stabilisation.

[Neurosurgery](https://rondamedtech.ro/en/specialitati/neurochirurgie/)

![BDyn fitted on one level, on a lumbar spine model](https://rondamedtech.ro/produse/cousin-surgery/bdyn-1-nivel.webp)

## Description

BDyn is a spinal shock absorber, a flexible spinal rod fixed posteriorly to pedicle screws, for the segments from T10 to S1. The device allows the anatomical movements of the segment: flexion, extension, axial rotation, compression and lateral bending.

Motion is controlled by two components: a silicone cushion that absorbs loads and helps decrease intradiscal pressure and facet loads, and a polycarbonate urethane (PCU) ring that controls hypermobility and limits extension. The upper rod is polyaxial at 14°, and the fixed lower rod is angled at 10° for lordosis.

Besides the one-level dynamic rod, BDyn has longer hybrid rods that combine fusion over one or more levels with dynamic stabilisation of the adjacent segment, in 3- or 5-level constructs.

During implantation, screw position is checked with the BDyn trial implant, which must sit flat against the base of the polyaxial heads without touching the facet joints. A 2 mm reserve, laser-marked on the rod, must be kept between the cylindrical body and the upper screw head so that the shock absorber works correctly; the construct is checked radiologically. For correct implantation, the authors of a 2023 study recommend intraoperative spinal navigation and at least 27 mm between the heads of two adjacent screws.

![BDyn hybrid construct: fusion with dynamic stabilisation of the adjacent segment](https://rondamedtech.ro/produse/cousin-surgery/bdyn-hibrid.webp)

## Videos

Film: https://rondamedtech.ro/video/cousin-surgery/bdyn-prezentare.mp4

BDyn system presentation (3D animation) · 1:19 · en

## Indications

-   Degenerative disease of the intervertebral disc and/or facet joints, confirmed by further examinations
-   Spinal canal stenosis
-   Grade 1 degenerative spondylolisthesis
-   Segmental instability
-   Prevention of adjacent disc degeneration, including in hybrid constructs above a fusion

## Features

-   Posterior dynamic stabilisation, from T10 to S1
-   Allows flexion, extension, axial rotation, compression and lateral bending
-   Silicone cushion that absorbs loads
-   PCU ring that controls hypermobility and limits extension
-   Polyaxial upper rod at 14°
-   Fixed lower rod angled at 10°, for lordosis
-   Hybrid rods for fusion combined with dynamic stabilisation of the adjacent segment

## Technical specifications

Technical specifications: BDyn

One-level dynamic rod

47 mm; Ø 5.5 mm (RCBDYSD55U)

Medium hybrid rod (3 levels)

107 mm; Ø 5.5 mm (RCBDYMD55U)

Long hybrid rod (5 levels)

230 mm; Ø 5.5 mm (RCBDYLD55U)

Fixed rod angle

10° for 1-2 level fusion; 0° for 1-4 level fusion

Upper rod

Polyaxial, 14°

Materials

Titanium, silicone, polycarbonate urethane (PCU)

Levels

T10-S1

Classification

Implant Class III, CE 1639 (SGS Belgium NV); instruments Class Ir

Manufacturer

Cousin Biotech S.A.S., France

## Gallery

![BDyn hybrid rods](https://rondamedtech.ro/produse/cousin-surgery/bdyn-tije.webp)

![BDyn construct with a long rod, over 5 levels](https://rondamedtech.ro/produse/cousin-surgery/bdyn-hibrid-5-nivele.webp)

## Clinical studies and publications

1.  Surgical tips for the implantation of BDYN dynamic stabilization system: combining spinal navigation with motion preservation for low-grade lumbar degenerative spondylolisthesisSerratrice N, Tarabay B, Faddoul J, Abi Lahoud GN · _Acta Neurochirurgica 165(9):2717-2721_, 2023[DOI 10.1007/s00701-023-05716-1](https://doi.org/10.1007/s00701-023-05716-1)
2.  Radiological factors affecting functional outcome after the implantation of BDYN dynamic stabilization system for low-grade lumbar degenerative spondylolisthesisSerratrice N, Faddoul J, Tarabay B, Attieh C et al. · _Neurochirurgie 69(4):101456_, 2023[DOI 10.1016/j.neuchi.2023.101456](https://doi.org/10.1016/j.neuchi.2023.101456)
3.  Finite element modelling of hybrid stabilization systems for the human lumbar spineDemir E, Eltes P, Castro AP, Lacroix D, Toktas I · _Proc Inst Mech Eng H 234(12):1409-1420_, 2020[DOI 10.1177/0954411920946636](https://doi.org/10.1177/0954411920946636)
4.  Viscoelastic properties of a spinal posterior dynamic stabilisation deviceLawless BM, Barnes SC, Espino DM, Shepherd DE · _J Mech Behav Biomed Mater 59:519-526_, 2016[DOI 10.1016/j.jmbbm.2016.03.011](https://doi.org/10.1016/j.jmbbm.2016.03.011)
5.  In vitro oxidative degradation of a spinal posterior dynamic stabilization deviceLawless BM, Espino DM, Shepherd DE · _J Biomed Mater Res B Appl Biomater 106(3):1237-1244_, 2018[DOI 10.1002/jbm.b.33913](https://doi.org/10.1002/jbm.b.33913)

## Documents

-   [BDyn product sheetPDF · en · 231 KB](https://rondamedtech.ro/documente/cousin-surgery/bdyn-fisa-produs.pdf)
-   [BDyn surgical technique (2022)PDF · en · 692 KB](https://rondamedtech.ro/documente/cousin-surgery/bdyn-tehnica-chirurgicala.pdf)
-   [BDyn and IntraSPINE clinical book: study abstractsPDF · en · 685 KB](https://rondamedtech.ro/documente/cousin-surgery/bdyn-intraspine-studii-clinice.pdf)

## Frequently asked questions

### What movements does BDyn allow?

Flexion, extension, axial rotation, compression and lateral bending of the segment. The silicone cushion absorbs loads and helps reduce intradiscal pressure and facet loads, while the polycarbonate urethane ring controls hypermobility and limits extension.

### Can BDyn be combined with a fusion?

Yes. The hybrid rods, 107 mm for 3-level constructs and 230 mm for 5 levels, combine fusion over one or more levels with dynamic stabilisation of the adjacent segment. The fixed rod is angled at 10° for 1-2 level fusion and 0° for 1-4 level fusion.

### At which levels and for which indications is it used?

From T10 to S1, in degenerative disc and/or facet disease confirmed by further examinations, spinal canal stenosis, grade 1 degenerative spondylolisthesis and segmental instability.

### What sizes are the rods?

All rods are Ø 5.5 mm: the one-level dynamic rod is 47 mm, the medium hybrid rod 107 mm and the long one 230 mm. The upper rod is polyaxial at 14°. The implant is Class III.

## Quote request

Product

BDyn

Cousin Surgery

Fill in the form and we will get back to you shortly, usually within the same working day.

## Related products

Nexon Medical

### [EnSight Biportal](https://rondamedtech.ro/en/producatori/nexon-medical/ensight-biportal/)

A complete solution for unilateral biportal endoscopic spine surgery (UBE): 4K endoscopic tower, Plasma and RF electrodes, dedicated instruments and an endoscopic drill.

Product page

Nexon Medical

### [EnSight Drape Pack](https://rondamedtech.ro/en/producatori/nexon-medical/ensight-drape-pack/)

A complete sterile set for biportal and monoportal spinal endoscopy: drapes, camera cover, fluid collection and the small instruments needed, in a single pack.

Product page

Medacta

### [Mecta-C Stand Alone](https://rondamedtech.ro/en/producatori/medacta/mecta-c/)

A modular anterior cervical fusion (ACDF) system that combines a TiPEEK interbody cage with an anterior plate: 4 plate types, rigid or variable fixation, 8 configurations.

Product page

---
Quotes and information: https://rondamedtech.ro/en/contact/ · office@rondamedtech.ro · +40 727 221 112
